18-3
IM 765501-01E
Specifications
3
2
1
6
5
4
9
8
7
12
11
10
15
14
13
18
18
17
16
Index
App
18.2 Measurement Section
Voltage Measurement
Integration Time of 16.66 ms/20 ms, 100 ms, or 200 ms
Range
Measurement
Resolution
Accuracy (One Year)
Temperature Coefficient
Setting
Range
±
(% of V)
±
(% of V)/
°
C
200 mV
±
205.000 mV
1
µ
V
0.02+100
µ
V
0.002+30
µ
V
2 V
±
2.05000 V
10
µ
V
0.02+200
µ
V
0.002+20
µ
V
20 V
±
20.5000 V
100
µ
V
0.02+1 mV
0.002+100
µ
V
110 V
±
110.000 V
1 mV
0.02+10 mV
0.002+1 mV
Integration time of 4 ms, 1 ms, or 250
µ
s
Range
Measurement
Resolution
Accuracy (One Year)
Temperature Coefficient
Setting
Range
±
(% of V)
±
(% of V)/
°
C
200 mV
±
205.00 mV
10
µ
V
0.02+200
µ
V (300
µ
V)
0.002+20
µ
V (30
µ
V)
2 V
±
2.0500 V
100
µ
V
0.02+300
µ
V (500
µ
V)
0.002+30
µ
V (50
µ
V)
20 V
±
20.500 V
1 mV
0.02+3 mV (5 mV)
0.002+300
µ
V (500
µ
V)
110 V
±
110.00 V
10 mV
0.02+30 mV (50 mV)
0.002+3 mV (5 mV)
Accuracy: One year accuracy at 23
±
5
°
C with auto zero turned ON.
Temperature coefficient: Add the temperature coefficient at 5 to 18
°
C and 28 to 40
°
C.
Values inside the parentheses are those when those when the integration time 1 ms or 250
µ
s.
Current Measurement
Integration Time of 16.66 ms/20 ms, 100 ms, or 200 ms
Range
Measurement
Resolution
Accuracy (One Year)
Temperature Coefficient
Setting
Range
±
(% of A)
±
(% of A)/
°
C
20
µ
A
±
20.5000
µ
A
100 pA
0.03+50 nA
0.003+5 nA
200
µ
A
±
205.000
µ
A
1 nA
0.03+300 nA
0.003+30 nA
2 mA
±
2.05000 mA
10 nA
0.03+3
µ
A
0.003+300 nA
20 mA
±
20.5000 mA
100 nA
0.03+30
µ
A
0.003+3
µ
A
200 mA
±
205.000 mA
1
µ
A
0.03+300
µ
A
0.003+30
µ
A
3 A
±
3.20000 A
10
µ
A
0.03+5 mA
0.003+500
µ
A
Integration time of 4 ms, 1 ms, or 250
µ
s
Range
Measurement
Resolution
Accuracy (One Year)
Temperature Coefficient
Setting
Range
±
(% of A)
±
(% of A)/
°
C
20
µ
A
±
20.500
µ
A
1 nA
0.03+70 nA (80 nA)
0.003+7 nA (8 nA)
200
µ
A
±
205.00
µ
A
10 nA
0.03+350 nA (400 nA)
0.003+35 nA (40 nA)
2 mA
±
2.500 mA
100 nA
0.03+3.5
µ
A (4
µ
A)
0.003+350 nA (400 nA)
20 mA
±
20.500 mA
1
µ
A
0.03+35
µ
A (40
µ
A)
0.003+3.5
µ
A (4
µ
A)
200 mA
±
205.00 mA
10
µ
A
0.03+350
µ
A (400
µ
A)
0.003+35
µ
A (40
µ
A)
3 A
±
3.2000 mA
100
µ
A
0.03+5.5 mA (6 mA)
0.003+550
µ
A (600
µ
A)
Accuracy: One year accuracy at 23
±
5
°
C with auto zero turned ON.
Temperature coefficient: Add the temperature coefficient at 5 to 18
°
C and 28 to 40
°
C.
Values inside the parentheses are those when those when the integration time 1 ms or 250
µ
s.