App-10
IM 765501-01E
Procedure
Preparation
1.
Create an arbitrary sweep pattern file for varying base current I
B
of the transistor
using a general-purpose spreadsheet application on your PC.
2.
Save the data that you created in step 1 to a file in CSV format.
3.
Transfer (Copy) the pattern sweep data file to the internal memory
(GS610ROM) of the second GS610 via the USB.
4.
Set each GS610 as follows:
Setup Example of
Setup Example of
the First GS610
the Second GS610
SOURCE
Output function
Voltage output
Current output
Limiter value
150 mA
1 V
Output mode
DC
DC
Output range:
AUTO
AUTO
OUTPUT CONTROL
Trigger mode
External
Internal
Sweep mode
OFF
Program
MEASURE
Measurement function
ON
OFF
Measurement function
Current measurement
–
Integration time
200 ms
–
Measurement range
AUTO
–
Store count
Source data count
–
TIME
Source delay
1
µ
s
1
µ
s
Measurement delay
1 ms
1 ms
Repeat period
–
500 ms
(of the internal trigger)
External I/O
Trigger input
BNC input
–
Trigger output
–
BNC output
Appendix 2 Application Examples