10.2 Synchronized Operation
Multiple GS610s can be connected and operated in sync by using the input/output signal
function.
On the GS610, there is no configuration for setting master and slave units. However, the
words master and slave are used in the explanations.
Typical synchronization operations are shown below.
Synchronizing the Pulse Generation and Measurement
• Trigger mode setting
Master: Internal trigger (Int)
Slave:
External trigger (Ext)
For the setup procedure, see section 4.1, “Setting the Trigger.”
• Setting the Input/Output Signals of the BNC Input/Output Terminal and External
Input/Output Connector
Master: Set BNC Out to Trig and set Type to Origin or set DIO 6pin to Origin.
Slave:
Set BNC In to Trig.
For the setup procedure, see “Setting the BNC Input Terminal,” “Setting the BNC
Output Terminal,” or “Setting the External Input/Output Connector” in the previous
section.
• Connecting the Instruments
Connect Trig Out of the master unit to Trig In of the slave unit using the BNC input/
output terminal or external input/output connector.
For the pin assignments, see “Signal Name and Function” in the previous section.
AVERAGE
4 WIRE
KEY LOCK
DISPLAY
ESC
MEASURE
STORE
AUTO
RANGE
MENU
REGALL
MATH
NULL
COMPARE
MEASURE
VALUE
REMOTE
LOCAL
POWER
MISC
TIME
KEY LOCK
SHIFT
MODE
AUTO
RANGE
LIMIT
VS IS
MENU
SOURCE
OUTPUT CONTROL
SWEEP
START
ZERO
SOURCE
TRIG
MODE
OUTPUT
BS
V I
SENSE
OUTPUT
Hi
110V
MAX
1V
MAX
110V
MAX
Lo
7
8
9
4
5
6
1
2
3
0
.
+ -
RANGE
ALL TERMINALS
250V MAX TO
SOURCE MEASURE UNIT
GS610
AVERAGE
4 WIRE
KEY LOCK
DISPLAY
ESC
MEASURE
STORE
AUTO
RANGE
MENU
REGALL
MATH
NULL
COMPARE
MEASURE
VALUE
REMOTE
LOCAL
POWER
MISC
TIME
KEY LOCK
SHIFT
MODE
AUTO
RANGE
LIMIT
VS IS
MENU
SOURCE
OUTPUT CONTROL
SWEEP
START
ZERO
SOURCE
TRIG
MODE
OUTPUT
BS
V I
SENSE
OUTPUT
Hi
110V
MAX
1V
MAX
110V
MAX
Lo
7
8
9
4
5
6
1
2
3
0
.
+ -
RANGE
ALL TERMINALS
250V MAX TO
SOURCE MEASURE UNIT
GS610
Trig Out
(Origin)
Trig In
Master
Slave
Trigger setting: Ext
Trigger setting: Int
Synchronizing the Sweep Operation
• Trigger mode setting
Master: Internal trigger (Int)
Slave:
External trigger (Ext)
For the setup procedure, see section 4.1, “Setting the Trigger.”
• Setting the Input/Output Signals of the BNC Input/Output Terminal and External
Input/Output Connector
Master: Set BNC Out to Trig and set Type to Origin or set DIO 6pin to Origin.
Set DIO 7pin to Origin.
Slave:
Set BNC In to Trig.
For the setup procedure, see “Setting the BNC Input Terminal,” “Setting the BNC
Output Terminal,” or “Setting the External Input/Output Connector” in the previous
section.