Turning the Output ON/OFF and Generating Zero Output
(See sections 5.7 and 5.8 for the procedure)
As shown in the figure below, there are three output modes, Output ON, Output OFF,
and Zero Generation. The mode is switched with the OUTPUT and ZERO keys.
Output
ON
Output
OFF
Zero
output
OUTPUT
illuminates
ZERO
illuminates
SHIFT
ZERO
ZERO
ZERO
OUTPUT
OUTPUT
OUTPUT
Output OFF:
The output relay is OFF, and voltage or current is not generated.
The source measure cycle is also stopped.
Output ON:
The output relay is ON, and voltage or current is generated. The
source measure cycle is repeated.
Zero Generation: The output relay is ON. If generating voltage, 0 V is generated. If
generating current, 0 A is generated. Then, the source measure
cycle is stopped.
Zero Source Offset
An offset can be specified on the zero value when generating zero output.
Zero Source Impedance
Low impedance (LoZ) is default when generating voltage, and high impedance (HiZ) is
default when generating current. Only the source value is set to zero.
If high impedance is set when generating voltage or low impedance is set when
generating current, the source value is set to zero, and the limiter is constrained to the
minimum value of the range.
Offset Calibration
(See section 5.9 for the procedure)
such as due to temperature change. For example, the offset calibration is used to
correct the offset fluctuation due to the leak current of the internal circuit when using a
low current source range in a high temperature environment.
2.3 Source