XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
83
-
P/N: 9182
-
0189 Revision Q
Alert
Definition
Action needed
O2 cell not system
-
compatible
The incorrect type of oxygen
cell has been installed
Install an oxygen cell intended for
the HAMILTON C
-
2 (PN 396200)
Pressure Limitation
Inspiratory pressure, including
PEEP/CPAP, is 10 cmH2O below
Pressure. The ventilator limits
applied pressure, so the target
pressure or volume may not be
achieved
Check the lung for adequate ventila-
tion
Check ventilator settings and alarm
limits
NOTE: A cold lung will alert for
pressure limits until it warms up and
becomes more compliant
Preventative
maintenance re-
quired
According to its operating
hours, the ventilator requires
preventative maintenance
Have ventilator serviced
Replace HEPA filter
The air inlet HEPA filter shows
increased resistance
Replace the HEPA filter
Real time clock
failure
Date and time not set
Set date and time
Technical event:
XXXXX
A hardware or software mal-
function was detected. A tech-
nical alarm cannot typically be
corrected by the operator.
Ventilation will continue
Have the ventilator serviced
Turn Flow Sensor
The Flow Sensor connections
are reversed. Ventilation con-
tinues, but the ventilator cor-
rects for the reversed signal
Reverse the ends of the Flow Sensor.
The blue sensing line is close to the
lung and must be attached to the
blue connector. The colorless sens-
ing line is close to the ventilator and
must be attached to the white con-
nector.