) Instructions for Use
-
53
-
P/N: 9182
-
0189 Revision Q
4.4.1 High PA Pressures
Check position and patency of
the cannulas (Sec 4.4.2).
Confirm correct calibration of
sensors (Sec 3.3.8).
Re
-
check the flow meter
whenever a rise in PA Pressure is
detected and reduce perfusate
flow accordingly via reduction in
rpm settings.
NOTE: High peripheral vascular
resistance (PVR) is anticipated in
the re
-
warming period due to
vasoconstriction of the pulmo-
nary vasculature secondary to
hypothermia of the lung and the
perfusate. Strict control of the
PAP and a low flow modality (150
mL/min) will warm up the vascu-
lar tree without endothelial dam-
age and progressively lower PVR.
Troubleshooting
4.4 TROUBLESHOOTING