5.1 Connection, System
Limits and Date/Time
The XPS Software will generate
a Notice if a piece of equipment
is not connected to the main
touchscreen computer. This can
be due to lack of power to the
unit, a malfunction of the unit or
a physical cable disruption be-
tween the unit and the
touchscreen computer.
Select the NOTIFICATION
header to open up the equipment
window to see which piece of
equipment is causing the XPS
Software to generate a notice.
Another helpful tool is to press
the SERVICE button on the XPS
touchscreen. This will open the
Service window which shows each
piece of equipment and whether
or not it is connected to the
touchscreen computer.
A Date/Time Notice will gener-
ate if there is a >5 minute
difference between the XPS
touchscreen computer and any
other piece of hardware. Fix this
notice by resetting the date and/
or time of the offending hardware
item.
In the SERVICE page, adminis-
trators enter their User Name and
Password to access and adjust the
System Limits (range) of the PGM,
Ventilator, Pump and Graphs val-
ues by pressing the Update Limits
button. The High and Low Limit
values will control the color
scheme of the data values dis-
played on the Main page (green
= within limits, red = out of
range).
System Limits (Low and High)
can be configured for each pa-
rameter listed. The Save button
saves any changes, the Reset De-
faults button resets to the default
values, and the Close button clos-
es the page without saving
changes.
Priming: Notice