XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
20
-
P/N: 9182
-
0189 Revision Q
Remove 6 to 8 bottles of
STEEN Solution
TM
to allow to
warm up to room temperature.
Note: As a precursor to the rec-
ommended STEEN Solution
TM
Exchange step (Section 4.3.7)
consider warming half the bot-
tles (3 to 4 bottles) to a normo-
thermic temperature.
Ensure XPS is plugged into
the wall outlet. Note: wall outlet
should be equipped with back-
up power.
Turn toggle switch to the up
position to initiate XPS system
battery backup and press the
UPS button.
Start CardioHelp and HCU by
pressing power buttons for
each. Set HCU to 23 degrees.
Start Touchscreen monitor
(power button is located on the
back of the monitor), and Ven-
tilator by pressing power but-
tons for each.
2.3 Power Up
CAUTION: Before starting proce-
dure, ensure that all four wheel
casters are in locked position by
pressing each caster lock to the
DOWN position.