XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
76
-
P/N: 9182
-
0189 Revision Q
NOTE: The centrifugal pump
alerts listed on this page indicate
Medium
-
priority (
flashing
yellow)
alarms which can lead to alarm
status and eventual pump
stoppage if condition does not
improve.
Alert
Definition
Action needed
AC voltage error
There is an error reading from
the AC voltage sensor
Check wall power/ try another outlet
Contact XVIVO for service
Battery Defect
Battery 1 Defect
Battery 2 Defect
Battery very low
There is a problem with the DC
battery source power
Contact XVIVO for service
Bubble sensor defect
The bubble sensor has a prob-
lem reading data
Clean bubble sensor
Reset bubble detector
Device defect (Ox$)
There is an uncategorized de-
fect with centrifugal pump
Contact XVIVO for service
Device temp. too
high
The centrifugal pump internal
temperature is >42
o
C
Check room temperature
Check cooling fan operation
Contact XVIVO for service
Drive temp. not
available
Internal temperature sensor is
not working appropriately
Contact XVIVO for service
Ext. DC voltage error
The line voltage for the DC pow-
er source is out of range (11
-
28V)
Contact XVIVO for service
Level sensor defect
The level sensor is not reading
data
Reset level sensor
Replace level sensor
Check electrical connections to panel
LPM above limit
The liters per minute flow of the
pump are greater than the user
-
defined limit
Should be operating in RPM mode
LPM below limit
The liters per minute flow of the
pump are less than the user
-
defined limit
Should be operating in RPM mode
LPM out of range
Measured data does not match
displayed data
Contact XVIVO for service
Mode switched (LPM
<
-
> RMP)
Flow control mode was
switched by user from LPM to
RPM or vice
-
versa
Should be operating in RPM mode
Negative LPM
detected
The liters per minute flow of the
pump are negative (i.e. backflow
present)
Should be operating in RPM mode
Check clamps to ensure flow direction
is antegrade
No disposable
detected
Quadrox
-
iR oxygenator/pump
head is not properly connected
Re
-
seat oxygenator into housing
6.2.1 CardioHelp
XVIVO
(Medium
-
priority)