XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
67
-
P/N: 9182
-
0189 Revision Q
Notice
Definition
Action needed
Ventilator Peak
Invalid Data
The Peak Pressure data connec-
tion from the ventilator is
corrupt
Use ventilator readings for Ppeak for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator Peak Out
of Range
The Peak Pressure data is
outside of normal limits
Use ventilator readings for Ppeak for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator PEEP In-
valid Data
The PEEP data connection from
the ventilator is corrupt
Use ventilator readings for PEEP for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator PEEP Out
of Range
The PEEP data is outside of
normal limits
Use ventilator readings for PEEP for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator Static
Compliance Invalid
Data
The Cstat data connection from
the ventilator is corrupt
Use ventilator readings for Cstat for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator Static
Compliance Out of
Range
The Cstat data is outside of
normal limits
Use ventilator readings for Cstat for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator VTE
Invalid Data
The Tidal Volume data
connection from the ventilator
is corrupt
Use ventilator readings for V
T
for rest
of procedure and ignore XPS software.
Contact XVIVO to schedule service
Ventilator VTE Out of
Range
The Tidal Volume data is
outside of normal limits
Use ventilator readings for V
T
for rest
of procedure and ignore XPS software.
Contact XVIVO to schedule service
5.2 XPS
TM
Software
Priming: Notice