XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
121
-
P/N: 9182
-
0189 Revision Q
10.3.4 Control Setting Ranges and Resolutions:
Setting
Range
Accuracy
Default settings
Apnea backup
On, Off
—
On
ETS (expiratory trigger
sensitivity)
5
-
80%
5%
25%
Flow (l/min)
2 to 80
10% or 1 l/min whichever
is greater
15
Flowtrigger (l/min)
APVcmv, (S)CMV, PCV
+: 1
to 20, Off
5 l/min
Off
Gender
Male, Female
—
Male
I:E
1:9 to 4:1
—
1:2
Loudness (alarm)
1 to 10
1
5
% MinVol (% minute
volume)
25 to 350%
5%
100%
Mode
(S)CMV+, APRV, ASV
(Note: these are recom-
mended modes for lung ex
vivo ventilation)
—
(S)CMV+ (also called APV)
Oxygen
21
-
100%
± (volume fraction of 2.5%
+ 2.5% gas level)
50%
Pasvlimit
5
-
60 cmH
2
0
5% or 1 cmH
2
0
30 cmH
2
0
Patient Height
30
-
250 cm (3
-
129 kg IBW)
2 cm
174 cm
Pcontrol (control pressure,
added to PEEP)
5
-
60 cmH
2
0
5% or 1 cmH
2
0
15 cmH
2
0
PEEP
0
-
35 cmH
2
0
5% or 1 cmH
2
0
5 cmH
2
0
P
-
ramp
0
-
2000 ms
10 ms
100 ms
P low (APRV)
0
-
35 cmH
2
0
5% or 1 cmH
2
0
5 cmH
2
0
Rate
4
-
80 b/min (S)CMV+
1 b/min
7 b/min
Sigh
On, Off
—
Off
T high (APRV)
0.1
-
40 s
0.01 s
Based on rate (IBW)
TI (inspiratory time)
0.1
-
12 s
0.01 s
Based on rate (IBW)
Tlow (APRV)
0.2
-
40 s
0.01 s
Based on IBW
Vt (tidal volume)
20
-
2000 ml
10% or 10 ml
Based on IBW
C3 Ventilator
10.3 Technical Specifications