XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
13
-
P/N: 9182
-
0189 Revision Q
Lift the blue chamber table
until it clicks into the locked
position. Place the weight sensor
on the table, within the groves
(see section 2.6 for more detail).
Connect the weight sensor
cable, at the vent panel. To dis-
connect cable, depress clips (not
shown) on each right and left side
of connector.
Open the separately wrapped
sterile clear
“
U
”
drape (product #
1019) and wrap the non
-
cut end
across the weight sensor and blue
chamber table. Wrap the end
with the pre
-
cut slot up and over
the top cross bar. Leave a small
opening from the pre
-
cut slot
near the area where the perfusion
tubing will pass into the sterile
area.
Secure in place with stainless
steel bar clips.
NOTE: Clips can be autoclaved or
drape can be clipped by perfu-
sionist from non
-
sterile side.
Open the XVIVO Organ Cham-
ber
™
outer box. Remove the
Chamber and open the plastic
wrap by tearing along the blue
dotted line that reads
“
TEAR TO
OPEN
”
beginning at notch on
either end.
Remove the Chamber from the
poly bag and prepare to pass it
off aseptically to the gowned
scrub tech/nurse. Carefully and
aseptically, unwrap the first blue
drape layer away from the Cham-
ber and pass off to a scrub tech
or nurse.
Securely place the Chamber
on the weight scale to open the
second layer, using it to create a
sterile field around the Organ
Chamber.
2.2.1 XVIVO Organ
Chamber
TM
Disposable Set
-
up
MANUAL TIP:
Throughout the
XPS IFU, you routinely see three
color
-
coded number circles:
Black
represents an instruction
only (no associated picture).
Red
and/or
blue
indicates an
instruction and a representative
picture reference.