The objective of the EMC declaration is to enable the responsible organization to decide whether the
XPS is suitable for its electromagnetic environment.
The essential performance characteristic of the XPS is to pump STEEN Solution
TM
in an ex vivo lung
circulation model, with the aid of a centrifugal pump.
In foreseeable electrical and mechanical interferences, the XPS works as intended within the stated
accuracies.
The XPS is intended for operation in the electromagnetic environment detailed below.
The customer or the user of the device should ensure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment–
guidance
RF emissions
according to CISPR11
Group 1
The XPS uses RF energy only for its internal function.
Therefore, its RF emissions are very low and not likely to
cause any interference in nearby electrical equipment.
RF emissions
according to CISPR11
Class A
The XPS is suitable for use in all establishments other
than domestic and those directly connected to the
public low
-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
according to IEC 61000
-
3
-
2
Not Applicable
Voltage fluctuations/ flicker
emissions
according to IEC 61000
-
3
-
3
Not Applicable
XVIVO Perfusion Cart
System
10.7
Electromagnetic Compatibility (EMC) Declarations (IEC 60601
-
1
-
2)