XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
125
-
P/N: 9182
-
0189 Revision Q
10.3.8 Other Technical Data
Parameter
Specification
Patient ideal body weight (determined
from Pat. Height setting)
3 to 139 kg (6.6 to 306 lb)
*(actual patient weight can be higher (300 kg)
Inspiratory pressure
0 to 60 cmH
2
0
Maximum limited pressure
60 cmH
2
0
Maximum working pressure
0 to 60 cmH
2
0 (a combination of PEEP and
Pinsp). Ensured through pressure limiting.
Maximum inspiratory flow
240 l/min (150 l/min with 100% O
2
)
Tidal volume/ target tidal volume
20 to 2000 ml
Minute volume capability
Up to 60 l/min
Minimum expiratory time
20% of cycle time; 0.2 s to 0.8 s
Inspiratory valve response time
<13 ms
Automatic expiratory base flow
4 to 20 l/min
For Flowtrigger
≤ 2
l/min: 4 l/min
For Flowtrigger > 2 l/min: 2 x Flowtrigger
Means of inspiratory triggering
Flow (Flowtrigger control setting)
Oxygen mixer accuracy
± volume fraction of 2.5% + 2.5% of actual
reading
Oxygen cell life
1 year or 5000 h nominal. Actual cell life
depends on operating environment. Operation
at higher temperatures or higher oxygen
concentrations shortens cell life.
Alarm loudness
50 to 65dB (A) at 1 m
Tests and special functions
Tightness test, oxygen cell calibration, Flow
Sensor calibration, 100% O
2
, manual breath,
inspiratory hold maneuver, nebulization (30
min, 8 l/min), communications interface, com-
pensation of breathing circuit resistance and
compliance.
C3 Ventilator
10.3 Technical Specifications