XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
123
-
P/N: 9182
-
0189 Revision Q
10.3.5 Monitored Parameter Ranges, Resolutions & Accuracies
Parameter
Range
Accuracy
Oxygen
Oxygen
18
-
105%
± volume fraction of 2.5%
+ 2.5% of gas level
Vent Status panel
Oxygen
21
-
40%
—
PEEP
0
-
8 cmH
2
0
—
MinVol
0
-
350% of normal minute
ventilation expressed in
l/min
—
Pinsp
0
-
50 cmH
2
0
—
RSB
10
-
400 1/(l*min)
—
% fSpont
100
-
0%
—
Other calculated and displayed parameters
IBW
3
-
139 kg
—
10.3.6 Real
-
time curves and loops
Parameter
Range
Real time curves
Volume (V)
0 to 3200 ml
Flow
-
300 to 300 l/min
Airway pressure (Paw)
-
10 to 80 cmH
2
0
Time
0 to 15 s
Loops
Pressure/Volume
X: 0 to 3200 ml; y:
-
10 to 80 cmH
2
0
Volume/Flow
X: 0 to 3200 ml; y:
-
300 to 300 l/min
Pressure/Flow
X:
-
300 to 300 l/min; y:
-
10 to 80 cmH
2
0
C3 Ventilator
10.3 Technical Specifications