6.1 Heater/Cooler
Fault lamp (lights during alert
or alarm status).
Status and error message
display.
Alert silence button (silence
alert beep for 10 minutes).
Alert
Definition
Acton Needed
TEMP.DIFF > 1
o
C
Disconnect and reconnect hose
couplings during operation
Press silence button to silence
WATER LEVEL?
Water level too low
Sensor defect
Fill up water to just below MAX
-
mark on level indicator
Contact XVIVO customer service if
defect is suspected
CHECK UNIT
-
>
CUSTOMER SERVICE
Water tank empty
Sensor breakage/short circuit
Fill up water to just below MAX
-
mark on level indicator
Contact XVIVO customer service if
defect is suspected
CHECK UNIT
-
> C
TEMPERATURE TOO
LOW
Unit is too cold (<15
o
C)
Sensor breakage
Warm up unit for some time at
room temperature
Contact XVIVO customer service if
defect is suspected
22.0
27.0
WATER LEV-