XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
79
-
P/N: 9182
-
0189 Revision Q
6.2.2 CardioHelp
XVIVO
(Low
-
priority)
NOTE: The centrifugal pump
alerts listed on this page indicate
Low
-
priority (static yellow) alarms
which will not cause pump stop-
page, but need to be evaluated/
addressed.
Alert
Definition
Action Needed
LPM above limit
The LPM reading is over the
user
-
defined limit
Should be in RPM mode
LPM below limit
The LPM reading is below the
user
-
defined limit
Should be in RPM mode
No flow signal
The flow sensor is not connect-
ed
Reconnect flow sensor
pVen above warn limit
The venous (LA) pressure read-
ing is over the user
-
defined
limit
Check for kinks in lines
Reflush/zero pressure lines
Change alarm settings
pVen below warn limit
The venous (LA) pressure read-
ing is under the user
-
defined
limit
Check for kinks in lines
Reflush/zero pressure lines
Change alarm settings
pArt above warn limit
The arterial (PA) pressure read-
ing is over the user
-
defined
limit
Check for kinks in lines
Reflush/zero pressure lines
Change alarm settings
pArt below warn limit
The arterial (PA) pressure read-
ing is below the user
-
defined
Check for kinks in lines
Reflush/zero pressure lines
RPM above limit
The RPM reading is over the
user
-
defined limit
Check perfusion lines for kinks
Ensure clamps are positioned
correctly
Change RPM limit settings
RPM below limit
The RPM reading is below the
user
-
defined limit
Chang RPM limit settings
Software error
There is a software error with
the pump system
Watch for error to reset itself
Restart pump if possible
Switched to battery supply
The pump is running off of DC
battery power
Check wall power
Change wall outlet switch
Contact XVIVO for service
TVen above limit
The temperature reading from
the LA (reservoir) is over the
user
-
defined limit
Check heater/cooler unit
Change alert settings
TVen below limit
The temperature reading from
the LA (reservoir) is under the
user
-
defined limit
Check heater/cooler unit
Change alert settings