XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
134
-
P/N: 9182
-
0189 Revision Q
XVIVO Perfusion Cart System
10.8
Guidance and Manufacturer
’
s Declaration–
Electromagnetic Immunity
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380
–
390
TETRA 400
Pulse
modulation
b)
18 Hz
1,8
0,3
27
450
430
–
470
GMRS 460,
FRS 460
FM
c)
± 5 kHz deviation
1 kHz sine
2
0,3
28
710
745
780
704
-
787
LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0,2
0,3
9
810
870
930
800
-
960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0,3
28
1 720
1 845
1 970
1 700
—
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b)
217 Hz
2
0,3
28
2 450
2 400
—
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0,3
28
5 240
5 500
5 785
5 100
—
5 800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0,2
0,3
9
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000
-
4
-
3.
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be worst case.
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment