) Instructions for Use
-
137
-
P/N: 9182
-
0189 Revision Q
XVIVO Perfusion Cart System
10.10
XPS I/O Connections
Equipotential Ground Connector:
The Equipotential Ground Connector is used to equalize the potentials between different metal parts of the various Medical Electrical (ME)
equipment which make up a ME system, or to reduce the differences of potential which can occur during operation between ME devices and
conductive parts of other objects. The Equipotential Ground Connector may be connected directly between ME devices, or to a common ground
bus bar of the electrical installation. Reference IEC 60601
-
1 for ME systems.
The table below displays the I/O ports name and component connections used with the XPS
™.
The Touchscreen and Display
-
only monitors figures display the I/O ports name and component connection used (and not used) with the XPS
™.
Equipment
Connection
Notes
Electrical I/O Panel
Temperature sensor ports
LA and PA temperature sensors
Pressure sensor ports
LA and PA pressure sensors
Flow/Bubble sensor port
perfusate flow measurement and bubble
sensor
Ventilator I/O Panel
Inhale/Exhale ports
Ventilator breathing circuit
Air Flow
Ventilator flow sensor
High Pressure O2
High pressure oxygen connection (41
-
75 PSI)
Venous Gas
Venous gas connection from canister
Weight Sensor
Data communication
Heater/Cooler I/O Panel
Water lines
Connection to heat exchanger connections
on oxygenator
Venous Gas
Venous gas connection to oxygenator