S E R V I C E A N D G E N E R A L M A I N T E N A N C E
6-7
UV Filter Cap Kit Installation
(Excludes Micro-Spot instrument)
To install UV filter optics cap:
1.
With instrument resting on the top cover, rotate
optics cap counter-clockwise until the triangle on the
cap meets the diamond on the bottom housing.
Carefully remove the optics cap by lifting upwards
and set aside.
2.
Install UV filter optics cap by aligning triangle in cap
with the diamond in the bottom housing. Rotate
optics cap clockwise until the triangle on the cap
meets the triangle/circle on the bottom housing.
3.
Calibrate the instrument
. Refer to Section Three.
Diamond Mark
Triangle/Circle
Mark
Diamond Mark
Triangle Mark
Optics Cap