4
6.
Press the Enter key
#
to open the
Full Calibration
screen. <
Measure White
> patch appears in the User
dialog.
7.
Position the instrument on the reference white patch (as
explained earlier) and measure patch. Release the
instrument after <
Completed
> appears in the User
dialog.
8.
After the instrument is released,
Measure White
Again
appears in the User dialog. Measure white patch
one more time. Release instrument when <
Completed
>
appears.
9.
Make sure
Black
is highlighted and position instrument
on a black trap or measure open port.
10.
To measure open port (no trap),
unlatch the shoe and
point the instrument towards a dark area that is shielded
from light (such as under a table or desk). Press the read
switch (located in front of the optics) with your finger.
Make sure your finger does not cross the light path during
the measurement.
To measure with a black trap,
position the instrument on
the trap and lower the instrument to the shoe and hold
steady.
The instrument will go through a series of four
measurements.
11.
When <
Completed
> appears in the User dialog,
release the read switch or instrument. Press the Main
Menu key
%
to return to the main menu.
F u l l C a l i b r a t i o n
W h i t e
B l a c k
M e a s u r e b l a c k
t r a p o r
o p e n p o r t
<
M e a s u r e B l a c k
>