C H A P T E R T H R E E
3
-
2
snugly with the target window opening centered over
the white ceramic disk.
2.
Lower the instrument to the shoe; hold steady until
the user dialog indicates the calibration is completed.
NOTE:
If an
Optics Change?
message appears during the white
calibration, select the
No
option and re-measure calibration
reference. If the instrument optics was changed, a full calibration
must be performed. Refer to the Full Calibration procedure that
follows.
3.
Store the calibration reference in a dry, dust free area,
away from direct exposure to light.
C A L I B R A T I O N
W h i t e
M e a s u r e w h i t e
p a t c h o n
c a l i b r a t i o n
r e f e r e n c e
<
M e a s u r e W h i t e
>
Calibration Reference
White Patch