S E R V I C E A N D G E N E R A L M A I N T E N A N C E
6-3
Battery Pack Replacement
To replace the battery pack:
1.
Rotate the shoe stop 90° and carefully turn the
instrument over. This allows the shoe to open
perpendicular to the instrument housing.
2.
Using your fingers, compress the two tabs on the
battery access cover and remove it.
3.
Remove old battery pack from instrument and discard
of properly.
4.
MAKE SURE THE LABEL IS VISIBLE ON THE
NEW BATTERY PACK.
Slide the battery pack into
the compartment with the battery contacts toward the
rear of the instrument. Press down on the battery
pack to install it properly.
5.
Re-install the battery access cover into the instrument
housing and rotate shoe stop to its down position.
Battery Pack
Tabs
Battery Access Cover
Rotate Stop 90°