12.4.3.3 Fitting compression fittings
Procedure
1. Select correct compression fittings for the size of Watson-Marlow interface tubing to be used.
!
WARNING
Compression fittings may leak if used with the wrong interface tubing. Only use
Watson-Marlow interface tubing with Watson-Marlow compression fittings.
Procedure
2. Detach desired compression fitting from hydraulic connector pack.
3. Cut end of tubing so that it is square. As illustrated in the images below.
Procedure
4. Slide connection collar onto tube.
5. Slide compression ring onto tube ensuring inner step is facing cut end. As illustrated in the images
below.
12 INSTALLATION: PART 3 (FLUID PATH)
81