4-116 Maintenance Procedures
5. Fit the mounting shaft insertion tool to the mounting shaft.
6. Slide the lamp mounting flange onto the shaft, taking care to align it
correctly.
7. Reinsert the shaft through the side of the source enclosure, and fit the
lamp mounting flange to the APPI source enclosure side.
8. Use the combined 2.5-mm Allen wrench and cone extraction tool to
tighten the four mounting-flange securing screws.
9. Remove the mounting shaft insertion tool from the mounting shaft.
10. Fit the PEEK insulator to the end of the mounting shaft.
11. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit
and tighten the two insulator retaining screws.
Caution:
To prevent damage to the mounting shaft O-rings, fit the
mounting shaft insertion tool to the mounting shaft before fitting
the shaft to the lamp mounting flange.
Caution:
Tighten the securing screws sequentially and by small
increments until they are all fully tight. Doing so ensures that the
lamp mounting flange is uniformly seated on the APPI source
enclosure side plate.
Mounting shaft
insertion tool
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...