4-82
Maintenance Procedures
3. Fit the UNF coupling to the new capillary.
Requirement
: Use a UNF coupling with no grooves, which is appropriate
for the IonSABRE II probe.
4. Use the needle-nose pliers to slide a new ferrule onto the capillary.
5. Insert the capillary in the PEEK union, and ensure that it is fully
seated.
6. Screw the UNF coupling into the PEEK union, finger-tight only.
7. Gently tug on the capillary, testing to ensure that it stays in place.
8. Use the 7-mm wrench to tighten the locknut against the PEEK union.
9. Perform a leak test by attaching the free end of the PEEK tubing to an
LC pump and pumping 50:50 acetonitrile/water through it at 1 mL/min.
•
If leakage occurs, disassemble and remake the connection, and
repeat the leak test.
•
If the backpressure on the LC pump is high, replace the capillary,
and repeat the leak test.
10. When no leakage occurs and the backpressure on the LC pump is
normal, disconnect the PEEK tubing from the LC pump.
11. Remove the probe inlet connector and PEEK tubing from the PEEK
union.
12. Remove the probe heater (see
).
13. Fit the PEEK union/UNF coupling assembly to the nebulizer adjuster
knob.
14. Carefully thread the capillary through the probe assembly.
Warning:
To avoid eye injury from high-pressure liquid jet spray,
wear safety goggles when performing the leak test.
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...