4-84
Maintenance Procedures
Cleaning or replacing the corona pin
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
HPLC-grade (or better) methanol
•
Lint-free tissue
•
Lapping film
•
Corona pin
To clean or replace the corona pin:
1. Remove the corona pin from the source (see
) and inspect the
pin for damage.
2. Replace the corona pin if it is damaged; otherwise clean its tip with the
lapping film and a methanol-saturated tissue.
3. Install the corona pin in the source (see
Warning:
The probe and source components can be contaminated
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The probe and source can be hot. To avoid burn injuries, take
great care while performing this procedure.
Warning:
To avoid electric shock, ensure that the instrument is in
Standby mode before commencing this procedure.
Warning:
The corona pin tip is sharp. To avoid puncture wounds, handle
the corona pin with care.