Connecting the workstation
B-19
Connecting the workstation
Before connecting the workstation to the instrument, set up the workstation
according to its accompanying instructions. Locate the workstation within
5 meters (16 feet) of the instrument.
Requirement:
Use shielded network cables with the instrument to ensure
compliance with FCC, and other, limits.
To connect the workstation:
1. Connect the monitor to the PC.
2. Connect one end of the shielded network cable to the appropriate port on
the rear panel of the instrument.
3. Connect the other end of the shielded network cable to the port labeled
Instrument LAN on the workstation rear panel.
To connect the instrument to the power source:
1. Select the correct power cord for your location.
2. Connect the female end of the power cord to the power port on the rear
panel of the instrument.
Connecting Ethernet cables
Requirement:
Use shielded Ethernet cables with the instrument to ensure
compliance with FCC, and other, limits.
To make Ethernet connections:
1. Connect one end of one shielded Ethernet cable to the network switch,
and then connect the other end to the Ethernet card on the
preconfigured ACQUITY™ workstation.
Tip:
On preconfigured systems, the Ethernet card is identified as the
Instrument LAN card.
Caution:
Do not connect the instrument’s power supply cord until you
complete the installation procedures in the previous sections.
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...