Cleaning the sampling cone assembly
4-29
2. Grasp the cone gas nozzle handle, and use it to rotate the sampling cone
assembly 90 degrees, moving the handle from the vertical to the
horizontal position.
3. Slide the sampling cone assembly out of the ion block assembly.
Caution:
Do not open the isolation valve if the sampling cone
assembly is not attached to the ion block assembly.
TP03131
Cone gas nozzle
handle
Sampling cone
assembly
TP03132
Ion block assembly