4-70
Maintenance Procedures
3. Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the 3 screws retaining probe end-cover.
4. Remove the end cover and gasket from the probe assembly.
5. Unscrew and remove the nebulizer adjuster knob.
End-cover retaining screws
Nebulizer adjuster knob
End cover
Gasket
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...