ESI mode
3-3
3. With the probe label facing you, carefully slide the ESI probe into the
hole in the probe adjuster assembly, ensuring that the probe location
dowel aligns with the location hole of the probe adjuster assembly.
4. Tighten the probe locking ring to secure the probe in place.
Tip:
An automatic pressure test runs when the probe is correctly seated.
5. Connect the ESI probe’s cable to the high-voltage connector.
Caution:
To avoid failure of the automatic pressure test, fully
tighten the probe locking ring.
TP03129
Location hole of the probe
adjuster assembly
Probe location dowel
Probe label
Probe locking ring