Cleaning the ion guide assembly
4-55
18. Hold the sampling cone assembly so that the cone gas nozzle handle is
oriented horizontally and at the top, and then slide the sampling cone
assembly into the ion block assembly.
19. Grasp the sampling cone assembly handle, and use it to rotate the
sampling cone assembly through 90 degrees.
20. Fit the ion block assembly to the source assembly (see
Cleaning the ion guide assembly
Clean the ion guide assembly if cleaning the ion block and isolation valve fails
to increase signal sensitivity.
Removing the ion block assembly and ion guide from the source
assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
3-mm Allen wrench
•
O-ring removal kit
To remove the ion block assembly and ion guide:
1. Remove the source enclosure from the instrument (see
2. Remove the ion block assembly from the PEEK ion block support (see
).
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear chemical
resistant, powder-free gloves while performing this procedure.
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...