B-20 External Connections
2. Connect one end of the other shielded Ethernet cable to the back of the
instrument, and then connect the other end to the network switch.
I/O signal connectors
The instrument’s rear panel includes two removable connectors that hold the
screw terminals for I/O signals. These connectors are keyed so that they can
receive a signal cable inserted only one way.
Warning:
To avoid electric shock, separate all electrical connections to
the rear panel from hazardous voltages by double or reinforced
insulation. Circuits of this type are classified as safety extra low voltage
(SELV). Examples of circuits that are typically SELV include contact
closure inputs and outputs for autosamplers, and UV, RI, and
fluorescence detector signal outputs for LC/MS systems. The electrical
connections on the rear panel of this mass spectrometer are all SELV.
To avoid electric shock and damage to the instrument, do not apply more
than
• ±30 V d.c. to the Analog (Out) connection.
• 30 V d.c. to the Stop Flow (Out), Inject Start (In), Switch 2 (Out),
Switch 3 (Out), and Switch 4 (Out) connections.
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...