Combined APPI/APCI source
3-15
Removing the IonSABRE II probe and APPI/APCI source
enclosure
Required material
Chemical-resistant, powder-free gloves
To remove the combined APPI/APCI source
:
1. Prepare the instrument for working on the source (see
.
2. Remove the IonSABRE II probe (see
).
3. Disconnect the HT cable from the instrument’s front panel.
4. Disconnect the APPI drive cable from the instrument’s front panel.
5. Remove the source enclosure (see
6. Remove the corona pin (see
7. Fit the blanking plug to the pin’s mounting contact.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, ensure that the instrument is
prepared for working on the source before commencing this procedure.
Warning:
The probe and source can be hot. To avoid burn injuries,
take great care while working with these components.