3.6.2 Installing ionKey source software
If you are installing an ionKey source on your Xevo TQ-XS for the first time, you must install the
appropriate MassLynx software SCN and the ACQUITY UPLC M-Class driver pack. For further
details, see the following documents:
•
ACQUITY UPLC M-Class System Guide
(part number 715003588) for detailed installation
procedures, and information on using the ACQUITY Inlet Switch Utility.
• MassLynx software v4.2 and related SCN release notes for detailed information about
installing MassLynx software and SCNs.
3.6.3 Installing the camera in the ionKey source
To install the camera in the ionKey source:
1.
Connect the camera cable from the video output connector on the mass spectrometer’s
rear panel to the video-to-USB converter box.
Notice:
To avoid damaging the video converter, make sure the workstation is
powered-off before connecting the converter to the workstation in the next step.
2.
Connect the video-to-USB converter box to a USB port on the mass spectrometer’s
workstation.
3.
On the Tune page, click
Camera Viewer
.
4.
In Device settings dialog box, specify the parameter settings according to the following
table, and then click
OK
.
Tip:
After you install the camera software, when you select the ionKey camera viewer for
the first time, the device settings dialog box opens. To subsequently open the device
settings dialog box, in the camera viewer, click
View
>
Camera Options
.
Table 3–3: Device settings for the camera:
Parameter
Setting
Video norm
PAL_B
Video format
Y800 (768 x 576)
Frame rate (FPS)
25
Input channel
00 Video: Composite
3.6.4 Removing an ionKey source
You can remove the ionKey source, and replace it with a conventional interface.
June 9, 2016, 715004990 Rev. A