Required materials
• Chemical-resistant, powder-free gloves
• Waste container
To connect the liquid waste line:
Warning:
To avoid personal contamination with biohazards, toxic materials, and corrosive
materials, wear chemical-resistant gloves when performing this procedure.
1.
Place a suitable waste container below the mass spectrometer.
Notice:
To avoid distorting the drip tray or causing the drain cup to leak, restrain
the drain cup when attaching or removing the waste line.
2.
Slide a drain line over the barbed fitting of the drain (located at the bottom, left-hand side of
the mass spectrometer, and on the base of the bottle tray).
Figure B–8:
Drain cup
Barbed fitting
Drain line
June 9, 2016, 715004990 Rev. A
Page 202