Notice:
Do not rotate the impactor pin when wiping it in the lapping film or when
wiping the pin clean, because this can damage the pin.
4.
Install the impactor pin in the UniSpray source (see
Removing and installing the UniSpray
4.29 Replacing the fluidic lines of the ionKey source
If a fluid line becomes damaged, replace it to ensure optimal performance. Each fluid line is pre-
fitted with a PEEK connector assembly at the point where it connects to the ionKey source
module. You must replace the capillary tubing and the PEEK assembly as a single unit. No parts
of the fluid line are reusable once detached.
Figure 4–67:
PEEK fluid connector
Capillary tubing
O-ring
Compression fitting
Connector plate
To identify the correct tubing assembly for each fluid line, use the following table.
Part number
Order number
Description
430004188
700010399
Inlet tube
430004190
700010400
Infusion tube
430004212
700010401
Waste tube
430004476
700010470
Optional, post-column addition tube