Figure 4–26: StepWave assembly removal and insertion tool
Ion block support screw holes (2)
Brown PEEK ion guide cap
Slot
Cutout
Pins
Notice:
To avoid damage when removing the StepWave ion guide assembly
from the adaptor housing, handle only the brown PEEK ion guide cap
4.
Inserting your thumbs through the slots in the StepWave removal and insertion tool, pull
the StepWave ion guide assembly from the pumping block assembly and into the
StepWave removal and insertion tool.
Notice:
To avoid damaging the StepWave ion guide assembly, handle it and its
components carefully throughout the cleaning procedure.
In particular, to avoid damaging the wiring on assemblies fitted with an externally
wired printed circuit board (PCB), do not touch the wiring.
5.
Remove the StepWave ion guide assembly from the StepWave removal and insertion tool.
June 9, 2016, 715004990 Rev. A