1.
Fit the cone gas nozzle handle to the cone gas nozzle and turn the handle clockwise to
tighten.
Figure 4–13:
Cone gas nozzle handle
Cone gas nozzle
Sample cone
O-ring
2.
Carefully fit the sample cone into the cone gas nozzle.
3.
Fit the O-ring (a new one, if you disposed of the old O-ring) into the groove created
between the sample cone and cone gas nozzle.
4.13.5 Fitting the sampling cone assembly to the source
Required materials
Chemical-resistant, powder-free gloves
To fit the sampling cone assembly to the source:
Warning:
To avoid personal contamination with biohazards, toxic materials, and corrosive
materials, wear chemical-resistant gloves when performing this procedure.
Warning:
To avoid puncture injuries, handle sample needles, syringes, fused silica
lines, and borosilicate tips with extreme care.
Notice:
To avoid damage, do not open the source isolation valve before fitting the
sampling cone assembly to the ion block assembly