1.
Pull the source enclosure release (located at the bottom, right-hand side) outwards, and
swing open the enclosure.
2.
Using both hands, grasp the source enclosure, and lift it vertically off the two supporting
studs on the source adaptor housing.
3.
Slide the first ion guide assembly end of the StepWave assembly into the StepWave
removal and insertion tool.
Tip:
The StepWave assembly can only be inserted in the correct orientation in the
StepWave removal and insertion tool.
Figure 4–35: Sliding the StepWave assembly into the StepWave removal and
insertion tool
StepWave assembly
Pins (2)
StepWave assemby removal and insertion tool
Cutout
4.
With the StepWave removal and insertion tool’s cutout uppermost, insert the StepWave
removal and insertion tool’s pins into the ion block support screw holes above and below
the aperture in the pumping block assembly.
Notice:
To avoid damage when removing the StepWave ion guide assembly
from the adaptor housing, handle only the brown PEEK ion guide cap
June 9, 2016, 715004990 Rev. A
Page 135