![Waters Xevo TQ-XS Mass Overview And Maintenance Manual Download Page 76](http://html1.mh-extra.com/html/waters/xevo-tq-xs-mass/xevo-tq-xs-mass_overview-and-maintenance-manual_3666255076.webp)
Instrument source interface door
High-voltage cable socket
Probe adjuster cable socket (yellow)
7.
Connect the high-voltage cable to the high-voltage cable socket on the mass spectrometer.
8.
Connect the probe adjuster cable to the probe adjuster cable socket on the mass
spectrometer.
9.
Close the instrument's source interface door.
10. Screw the probe assembly's PEEK fitting into the LC flow or syringe pump.
3.4.2 Removing the UniSpray source
You can remove the UniSpray source, and replace it with another compatible interface.
Required materials
• Chemical-resistant, powder-free gloves
Warning:
To avoid personal contamination with biohazards, toxic materials, and corrosive
materials, wear chemical-resistant gloves when performing this procedure.
Warning:
To avoid electric shock, ensure that the instrument is prepared for working
on the source before commencing this procedure.
To remove the UniSpray source:
1.
Prepare the instrument for working on the source (see
Preparing the instrument for working
).
Warning:
To avoid burn injuries, exercise care when handling the components
of the source enclosure heated to high temperatures. Wait until the hot
components have sufficiently cooled before you handle them.
2.
Disconnect the PEEK fitting connecting the probe assembly to the LC.
3.
Swing open the UniSpray source enclosure unit from the source mounting on the mass
spectrometer.
4.
Disconnect the probe adjuster cable.
5.
Disconnect the high-voltage cable.
6.
Carefully remove the UniSpray source module, and store it safely.
June 9, 2016, 715004990 Rev. A
Page 76