Warning:
To avoid electric shock, do not insert any item into the probe cap aperture
when the probe cap is fitted to the instrument.
Notice:
To avoid damaging the probe assembly, take care when inserting the capillary
into the probe adaptor. Do not use force.
Warning:
To avoid harmless, static-like electric shock, ensure the mass spectrometer
is in Standby mode before you touch any of its external surfaces that are marked with
this high voltage warning symbol.
1.
Carefully insert the probe assembly capillary into the probe adaptor.
Tip:
To aid insertion, turn the capillary gently as you insert it, feeding the entire capillary in
to the probe adaptor.
Figure 3–12: Inserting the probe assembly
Probe assembly capillary
Probe adaptor
2.
Screw the probe adaptor fitting into the probe adaptor, finger-tight only, until you hear a
click.