During normal operation, the ionKey source does not require cleaning. After repeated use,
however, material can accumulate around the fluid-line connectors or electronic connectors in the
iKey docking port.
Figure 4–73:
Electronic connectors
Fluid line connectors
To remove buildup from fluid-line connectors:
Gently swab the fluid-line connectors and outer edges of the docking port with a lint-free tissue
saturated with isopropanol or methanol.
4.30.1 To remove buildup from electronic connectors
To remove buildup from fluid-line connectors:
1.
Open the source door.
2.
From the back, gently direct a stream of air from a clean source across the electronic
connectors.
Note:
To prevent contaminating the inside of the source, do not direct air into the source.
3.
Close the source door.
June 9, 2016, 715004990 Rev. A