To remove the ESI probe tip and gasket:
Warning:
To avoid personal contamination with biohazards, wear clean, chemical-
resistant, powder-free gloves when performing this procedure.
Warning:
To avoid burn injuries, take great care while performing this procedure.
Warning:
To avoid puncture injuries, handle sample needles, syringes, fused silica
lines, and borosilicate tips with extreme care.
1.
Remove the ESI probe adaptor from the source (see
).
2.
Use the 7-mm wrench in conjunction with the 10-mm wrench to remove the probe tip.
Figure 4–38:
10-mm wrench
Probe tip
ESI probe adaptor
7-mm wrench
3.
Remove the metal gasket from the probe tip.
June 9, 2016, 715004990 Rev. A
Page 140