![WATANABE WPMZ-1 Instruction Manual Download Page 76](http://html1.mh-extra.com/html/watanabe/wpmz-1/wpmz-1_instruction-manual_3499830076.webp)
Graphical Digital Panel Meter WPMZ-1 INSTRUCTION MANUAL
75
IM-0882-03
18.
TROUBLESHOOTING
ERROR MODE
Display on occurrence of an error
When some malfunctions occur, error codes are displayed according to the factor of the error.
List of error codes and recovery procedures
When some malfunction occurs, an error code is displayed according to the factor of the error.
ERROR
CODE
ERROR MESSAGE
RECOVERY PROCEDURE
E000
Program sum error
During the error mode, hold down the
ENTER key for 1 second (long-press) to
reset or power down and on.
*If the WPMZ does not recover by this
procedure, please contact your dealer or our
company.
E006
RAM error
E100 to 102
Errors associated with serial
flash memory
E103 to 105
E210 to 211
Errors associated with FRAM
E202 to 203
Errors associated with
calibration values
E110 to 111
Error associated with sensor
power short
E204 to 205
Errors associated with setting
values
E206 to 209
Errors associated with initial
values
Other than
above codes
Other errors
・
If start-up delay is enabled, the WPMZ displays “-----” according to the delay time.
・
If display value becomes out of displayable range, “OVER” is displayed in the display.
CAUTION
If error display is not recovered by system reset or power re-activation, please let us know the error
code and situation.
During error mode, outputs are disabled.