Graphical Digital Panel Meter WPMZ-1 INSTRUCTION MANUAL
38
IM-0882-03
Operation:[MENU]
⇒
[Input]
⇒
[DC InputA/B][ProcessInputA/B]
⇒
[InsDispStep]
Setting value: None (Default): 0 to 9
5steps
:
0 to 4 → 0,
5 to 9 → 5
10steps :
0 to 9 → 0
ZeroArea
Zero detection is a function to judge whether the display value is around zero and disable comparison output
function when it is near zero.
It is effective when measuring the weight of an object, such as suppressing comparison output function when not
holding anything on the tare.
In this item, set the range of the display value considered to be near zero.
* To enable this function, it is necessary to set the “ActCondition” of the comparison output function.
(The setting method of the “ActCondition” is [MENU]
⇒
[Output]
⇒
[CompareAL1
~
4]
⇒
[ActCondition]
Operation:[MENU]
⇒
[Input]
⇒
[DC InputA/B][ProcessInputA/B]
⇒
[ZeroArea]
Setting value: 0
~
99999 (Default is 0)
StableArea
It judges whether the display value is stable or not and invalidates the comparison output function when it is not
stable. If the variation of the display value at each specified time is within the set value, it is judged stable and
the “st” mark will light.
It is effective when measuring the weight of the object caught with a crane when you want to make a comparison
output function after the shake falls.
In this item, set the fluctuation range of the display value considered stable.
* Please set it together with “StableTime”. This function is invalid if either set value is 0.
* To enable this function, it is necessary to set the “ActCondition” of the comparison output function.
(The setting method of the “ActCondition” is [MENU]
⇒
[Output]
⇒
[CompareAL1
~
4]
⇒
[ActCondition]
Operation:[MENU]
⇒
[Input]
⇒
[DC InputA/B][ProcessInputA/B]
⇒
[StableArea]
Setting value: 0
~
99999 (Default is 0)
StableTime
Same function as “StableArea” and this item is a time parameter for detecting stability of display value.
* Please set it together with “StableArea”. This function is invalid if either set value is 0.
Operation:[MENU]
⇒
[Input]
⇒
[DC InputA/B][ProcessInputA/B]
⇒
[StableTime]
Setting value: 0
~
9999 * 10msec (Default is 0)
*You can set by 10ms steps.
2InputCalc
InsDispStep
When the minimum digit of display value fluctuates, truncate the minimum digit and stabilize the display.
Operation:[MENU]
⇒
[Input]
⇒
[2InputCalc]
⇒
[InsDispStep]
Setting value: None (Default): 0 to 9
5steps
:
0 to 4 → 0,
5 to 9 → 5
10steps :
0 to 9 → 0