![WATANABE WPMZ-1 Instruction Manual Download Page 28](http://html1.mh-extra.com/html/watanabe/wpmz-1/wpmz-1_instruction-manual_3499830028.webp)
Graphical Digital Panel Meter WPMZ-1 INSTRUCTION MANUAL
27
IM-0882-03
2 INPUT CALCULATION
*In 1ch products, this setting item is invisible.
Setting of 2 input calculation is proceed as follow.
Each setting items are set up at [Setting]
⇒
[Input]
⇒
[2InputCalc]
Descriptions of each setting item are described on the following pages after this section.
Moreover, please check "input detailed setting" for the setting not described in the above flow.
*1 Each setting and calibration value are saved and reflected at the timing when the menu is closed and
returned to
the measurement screen.
Note that the contents will not be saved if the power is turned off before closing the menu.
PatternSelect
You can store and set up 8 patterns (8 kinds) of various settings such as input and output with this unit, you can
select the pattern to save the setting in "PatternSelect".
The pattern number being set is displayed as "P5" on the upper right of the setting screen as shown below.
Operation:[MENU]
⇒
[Input]
⇒
[2InputCalc]
⇒
[PatternSelect]
Setting value:Pattern1,Pattern 2,
・・・・・
Pattern 8
* At the time of opening the menu, the setting pattern is the value selected as operating on the measurement
screen.
* Since the setting pattern is common in each setting, "PatternSelect" such as output setting is also changed
when "PatternSelect" is changed by "2InputCalc”.
Expression
Select arithmetic expression for 2 input signals.
*[A] means value of Ach, and [B] means value of Bch. The most detailed value is used for calculation.
(If display value is 123.45, calculation is done with 123.456789 that is the most detail internal data.)
Pattern select
Select arithmetic
expression
Select arithmetic
coefficient
Decimal point
position select
Unit select