![WATANABE WPMZ-1 Instruction Manual Download Page 35](http://html1.mh-extra.com/html/watanabe/wpmz-1/wpmz-1_instruction-manual_3499830035.webp)
Graphical Digital Panel Meter WPMZ-1 INSTRUCTION MANUAL
34
IM-0882-03
Threshold
Set threshold for comparison output.
Setting value of [Threshold] changes by setting value of [CompareMode].
Operation:[MENU]
⇒
[Output]
⇒
[CompareAL1], [CompareAL2], [CompareAL3], [CompareAL4]
⇒
[Threshold]
[CompareMode] is “Leveljudge”
Setting value: Threshold
±99999 (Default is10000)
Hysteresis 0
~
99999(Default is 0)
[CompareMode] is “Zonejudge”
Setting value:
ZoneLowerLimit
±99999 (Default is 0)
ZoneUpperLimit
±99999 (Default is 10000)
Hysteresis 0
~
99999(Default is 0)
*Same hysteresis is applied in ZoneLowerLimit and ZoneUpperLimit.
[CompareMode] is “Diffjudge”
Comparison output turns ON when
(
MAX - MIN
)
> ChangeAmount (See below)
Setting value: ChangeAmount
0
~
99999 (Default is 10000)
DetectionInterval 0.00
~
99.99[sec] (Default is 1.00)
*If “DetectionInterval” is 0.00, “Diffjudge” is disabled.
“Diffjudge” operating diagram
*If setting value of [OutputMode] is “Normal”, comparison output continues until end of “DetectionInterval”.
*When Compare Reset is ON, comparison output is always set to OFF.
【DetectionInterval】
【DetectionInterval】
Display value
OutputMode
(OneShot)
ON
MAX
MIN
MAX
MIN
(MAX-MIN)
(MAX-MIN)
OutputMode
(Latch)
ON
OFF
OFF
ON
OFF
OutputMode
(Normal)
Compare
Reset
Function ON