VET-PRO
Operator’s Manual
8
Specifications
Pumping Mechanism:
Piston Mechanism
Flow Rate:
PCA: 0 to 100 ml/hr in 0.1ml increments
All other programs: 0.1 to 1200 ml/hr in 0.1ml or
1 ml increments.
Total Infused Volume:
PCA: 0.1 to 1,000 ml.
Continuous, TPN, Intermittent: 0.1-9999
ml
25 Steps: 0.1 to 9999 ml. for each step
Accumulated volume for all steps- up to 10 liters.
Total Time Setting
100 hours
Accuracy:
±
3 %.
KVO rate:
0.1 to 5 ml/hr. (to be set)
Air Sensor:
Ultrasonic, adjustable air bubble size
Maximum Pressure:
0.7 bar or 10 psi Adjustable (high/normal/low)
Power Supply:
110-240 VAC, 50/60 Hz. 20 W
Battery:
Li-Ion 7.2V, 1450mA. (Rechargeable)
Battery Operation at 125 ml/hr 20 hours
Battery Charging:
Automatic when latched into the Charger
connected to an AC power source (4 hours)
Alarms:
End Program
Low & End Battery
Air in line
Down Occlusion
Door Open
Pump Unattended
Internal Malfunction
Locking state
Missing Key
Dimensions:
112 x 89 x 32mm. (L x w x h).
Symbol
Attention (Consult accompanying documents)