Operator’s Manual
39
I
Warning
Always disconnect IV tubing from patient before starting
flushing procedure.
4.
Select the “Modify Program” option and press
the
START/OK
key.
P
C
A
Select Program
=>
Modify Program
More
5.
Use the numeric key to set the code 008 and
press the
START/OK
key.
P
C
A
Enter Code
008
Press OK key
Epidural Mode
OFF
6.
Adjust ?? & press OK
If Epidural mode is required (see p.53) use
arrows key to change the setting to ON and
approve using the
START/OK
key.
To continue with epidural mode OFF, press
START/OK
key and move to step 8.
Air sensor alarm
ON
Adjust ?? & press OK
7.
If Epidural Mode was set ON, the user will
have to decide if to set the air alarm on or off.
Use arrows key to change the setting as
required and approve using the
START/OK
key.
8.
P
C
A
Clear History &
Start New Bag
Press the
START/OK
key to start new bag and
clear previous program data. Pump will display
bag volume screen - continue to step 9.
Pressing the
STOP/NO
key will display
program accumulation screen. Pressing
START/OK
key from accumulation screen
will take the VTBI left from previous program
as total vtbi for the new program. continue to
step 10.
9.
Pump will display the Bag Volume screen
(with total volume of previous program). Use
the numeric keys to set a new value or approve
the previous value, using the
START/OK
key.
P
C
A
Bag Vol: 500 ml
Press OK key
A:MORPHINE
B: METHADONE
10.
C: PETHIDINE
Pump will display the list of 5 protocols with
drug names. Use the arrow keys to scroll
through the options.
Pressing the
START/OK
key will display the
rate screen of the chosen protocol.
Pressing the
STOP/NO
key twice will enable
the user to change the drug name.